LC-1 Stand Alone Glovebox
Purpose
Creates an inert atmosphere for sensitive chemical experiments. It is capable of achieving and maintaining a <1ppm H2O and O2 environment.
Info
Make: LC Technology Solutions
Model: LC-1
Accessories: Mini Antechamber, Large Antechamber
Manual: http://remrsec-facilities.mines.edu/mediawiki/images/7/74/LC-1_Glovebox_Operation_Manual_rev6.pdf (Hard Copy also in lab)
Contact at Company: Tyler Kapsimalis, 978-255-1620
Location:36-362
Maintenance
Nitrogen Supply
Ni300 from Airgas with high purity.
Charcoal Regeneration
Requires forming gas (5% H, 95% N or Ar) with additional flow-indicating-regulator. New Tank can be purchased from supplier. Should be done every 3-4 months with heavy use, but less if not used often.
Solvent Trap Reactivation
Should be done when Oxygen levels start to regularly increase to over 50ppm. Requires only N gas.
Gloves
Should be changed when they start to show signs of erosion or wear
Rear White Filters
Should be changed when they start to yellow
1 Comment
Unknown User (eanquill@mit.edu)
Vac Pump currently undergoing maintennence